Sign In | Join Free | My enlightcorp.com |
|
Categories | Flammability Test Equipment |
---|---|
Brand Name: | GOLD |
Model Number: | GD-IEC 60112 |
Certification: | ISO |
Place of Origin: | China |
MOQ: | 1 set |
Price: | USD 3000-3500/set |
Payment Terms: | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Supply Ability: | 5 sets per month |
Delivery Time: | 30 days |
Packaging Details: | Plywood case |
Product name: | Tracking index tester |
Standard: | IEC 60112,ASTM D 3638,UL746A,JIS C2134 |
Application: | Insulation materials, plastics, etc. |
Test voltage: | 0-600V |
Electrode Angle: | 60°+-5° |
Sample pressure: | 1.00N+-0.05N |
Specification: | 630mmx390mmx750mm |
Dropping Time: | 30s ± 5s |
Company Info. |
Chongqing gold mechnical and electrical equipment Co., Ltd |
Verified Supplier |
View Contact Details |
Product List |
IEC 60112 PTI And CTI Comparative Tracking Indices Tester For Solid Insulating Materials
Tracking Indices Tester Introduction:
The tracking tester is a special testing instrument developed, designed and manufactured in strict accordance with GB4207, IEC60112 and other standards. It is suitable for research, production and quality inspection departments of lighting equipment, low-voltage electrical appliances, household appliances, machine tool electrical appliances, motors, power tools, electronic instruments, electrical instruments, information technology equipment, and is also suitable for insulating materials, engineering plastics, electrical connectors, and accessories industries. It simulates the determination of the tracking index and the tracking resistance index under humid conditions, and has the characteristics of simplicity, accuracy, reliability and practicality.
Tracking Indices Tester standard:
ASTM D 3638-92
IEC60112
UL746A
Tracking Indices Tester Test principle:
It is used to evaluate the tolerance of the surface of solid
insulating materials under the combined action of electric field
and contaminated medium, and to determine its comparative tracking
index (CTI) and proof tracking index (PTI);
Comparative Tracking Index (CTI): The highest voltage value that
the surface of the insulating material can withstand 50 drops of
electrolysis (0.1% ammonium chloride aqueous solution) without
forming a leakage trace, in V.
Proof Tracking Index (PTI): The voltage required for the material
surface to form a permanent conductive carbon path after
withstanding the action of 50 drops of electrolyte at a rate of one
drop per 30 seconds, expressed in V.
Tracking Indices Tester Sample requirements;
1. Plates and sheets with a diameter of 30 mm or more, or length
and width greater than 30 mm; thickness > 3 mm, samples can be
stacked, and the surface is flat;
2. Testable voltage: up to 600 V.
Tracking Indices Tester Main feature:
1.The combustion chamber paint treatment; nice,
corrosion-resistant; dimensions: 1200mm X 600mmx 1200mm;
2. All sample holder made of stainless steel manufacturing;
3. The device consists of voltmeter, ammeter, time overcurrent
relays, power supplies, circuit control system, push button
switches, lights and other components;
4. Automatic drip device includes dropping tube, drip nozzle,
counting means counting device drip, drip control device, dropping
the volume measuring tools;
5. Drip nozzle device can be adjusted; adjust the height of
30-40mm; ensure dripping position in the middle of the two
electrodes;
6. The electrode means comprises platinum electrode, the electrode
fixtures, 4mm gauge, etc., the electrode can be adjusted,
removable;
7. Electrode width 5mm, thickness 2mm, 99.9% purity platinum, 30 to
60 degrees tip;
8. The electrode sample pressure adjustable, adjustable range
(1.00N ± 0.001N); electrode distance 4.0mm ± 0.01mm, the angle 60 °
± 5 °;
9. With the sample stage;
10.With the glass: Thickness ≥4mm;
11. With electrolyte: sodium chloride and alkyl naphthalene
sulfonic acid reagent each bottle.
Safety control system:
1.The test voltage: 0 ~ 600V adjustable
The maximum error of voltmeter displays : 1%;
2. Short-circuit current: (1 Guests 0. 1) A,
The voltage drop value does not exceed 10%
3. The test current exceeds 0.5A duration of more than two seconds,
the system automatically shut down equipment;
4.Security door control system; the door opens, the test system is
not working;
5.The electrode: 5mm × 2mm rectangular platinum electrode and brass
electrode of each pair meet size requirements of IEC60112 Figure
B.1;
6.The electrode: the angle formed between the electrodes is 60 °
disabilities 5 °, a pitch of (4 Soil 0.1mm);
7.The electrode: the sample pressure: 1 00N disabilities 0.05N;.
8.Drip system: (30 ± 5) seconds (on time 29.8S + open drip drip
duration 0.2S) automatic counting, digital display (can be preset),
50 drops of time: (24.5 Soil 2) min;
9.Drip systems: a drop of fluid needle tip to the sample surface
height: 35mm disabilities 5mm (attach a gauge for measuring your
reference);
10.Drip systems: drip Weight: 20 drops: 0.380g ~ 0.489g; 50
drops:.. 0 997g ~ 1 147g;
11.PLC with touch screen control system; to achieve test smarter,
more convenient has been operated.
12.The test data is automatically recorded, automatic storage, free
retrieval;
Tracking Indices Tester Specification:
Electrode material | Electrode head - platinum (platinum) |
Electrode pressure | 1.0±0.05N |
Electrode distance | 4.0mm ± 0.01mm, angle 60°±5 |
Electrode voltage | 100~600V (48~60Hz) adjustable, voltage drop is less than 10% when short-circuit current is 1.0±0.1A |
Drop height | 30~40mm adjustable |
Drop size | 44~55 drops/1cm3 adjustable |
Drop time | 30s±5s |
Drop times | 0~999 times adjustable |
Leakage judgment | If the loop current is greater than 0.5A and remains for 2 seconds, the relay will cut off the current, indicating that the sample is unqualified |
Test process | The test procedure is automatically controlled, and the ventilation is independent; |
Reference standards | IEC 60112, GB/T 4207, GB 4706.1 |
Equipment dimensions | 1200mm wide × 600mm deep × 1300mm high |
![]() |